Orthopedic Multifunction Drill Saw for Veterinary Mini Surgical Bone Drill with Drill Bit Medical Oscillating Saw Power Tool
Specification
Parameter | Data |
Speed | 0-1200rpm |
Torque | 2.9N-m |
No-load Noise | ≤45dB |
Operation(V) voltage | 7.2V |
Sterilization temperature | 155° |
Radial circle run-out | ≤0.1mm |
Axial Movement | ≤0.5mm |
Chuck gripper diameter | ≤4mm |
Oscillations | 0-16000cpm |
Battery Capacity | Lithium 7.4V 2000mAh |
Accessories | Quantity |
Handpiece | 1(1.02kg,can be connected with several adapters,autoclavable up to 135℃) |
Battery | 2(can not be sterilized,could last for 30 minutes after fully charged) |
Charger | 1(110-220V,match both European & American type) |
Drill Chuck | 1(imported,high coaxiality,suitable for trauma surgeries) |
Oscillating Saw | 1(used for joint surgeries) |
Blades | 2(7 stars blades with optional sizes) |
Small AO | 1(trauma surgery for AO type,it's for drilling holes and fixing the plates) |
K-Wire | 1(holding K-wire very conveniently) |
Sterilizing channel | 1(could be sterilized) |
1. Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ), if the temperature is over 135 °C, the motor will be damaged.
2. Normal bone drill can only hold drill bits less than 6mm, acetabulum reaming drill with big torque,rpm is around 300, canulate drill is used with nail or K wire, rpm is around 600.
3. The battery need to be get out of the power tool after surgery. Otherwise the small automatic discharge of the electric will make the life of the battery short.
4. The battery cannot be sterilized, it is not autoclavable.
5. The cranial bur must be washed right after the surgery by purger in order to avoid the blood stick in the bur, then we should make it dry.
6. The battery space in the handpiece of the tool should not have any water,after sterilization.
7. The cranial bur blade is made of medical stainless steel, in order to avoid broken, so it is flexible and not very hard, it is easy to be attrite, we suggest each blade only use for 1 surgery.